Connecting
5.1 Electrical connections
5.1.2.2

Danger due to electrical current

Danger due to electrical current
The power cable to be connected at the customer location could be damaged by incorrect
fitting in the device.
You should therefore carry out a visual inspection and PE conductor test of the cable in
accordance with the country-specific safety standards, regulations or other applicable
directives before starting up the device. Because part of the cable is in the device, always
adhere to the product safety standard IEC/EN 61010-1 or its national equivalent.
5.1.2.3

Cable glands for the wall-mounted device

NOTICE
Cable glands for the wall-mounted device
Match the diameter of the cable to the threaded cable glands. After installation, make sure
that the cables are secure and that the seals are in the correct position.
5.1.3

Rack-mounted device

5.1.3.1

Connecting the power supply

Requirements
Note
Appliance plug
You can use any appliance plug that fulfills the conditions of DIN EN 60320-1, ≥ 6 A to
connect the device.
40
WARNING
Compact Operating Instructions, 06/2017, A5E31805656-05
Quick Start