Siemens SIPROCESS GA700 Quick Start Manual page 80

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Commissioning: Preparation
7.1 General safety instructions
See also
References (Page 108)
Note
Using an external pressure sensor
An external pressure sensor can process ambient or sample gas pressure values up to 3 bar
(abs.), regardless of the application.
You can obtain additional information on the use of external pressure sensors from your
service partner: → Technical support (Page 107).
Note
OXYMAT 7 pressure correction
The OXYMAT 7 analyzer modules have an internal pressure sensor for correcting the effect
of pressure on the measured value.
In the case of OXYMAT 7 the pressure sensor is connected with the sample gas path and
therefore always measures the current sample gas pressure.
Note
ULTRAMAT 7 pressure correction
If you are using the internal atmospheric pressure sensor, make sure that the pressure drop
caused by the piping system at the sample gas outlet does not exceed 10 hPa (large line
diameter, short line length). In analyzer cabinets and analyzer enclosure, you can connect
the pressure transducer to the atmosphere in order to record only the atmospheric pressure
change.
Note
CALOMAT 7 pressure correction
With CALOMAT 7, you can implement required pressure correction using an external
pressure sensor.
80
Compact Operating Instructions, 06/2017, A5E31805656-05
Quick Start

Advertisement

Table of Contents
loading

Table of Contents