Rectifying errors
If the device aborts a calibration due to errors, an error code, e.g., "F0123", and a plain text
error message are displayed in the result window.
1. Write down the error code.
2. If you cannot rectify the cause of the error yourself, forward the error code to the service
technician in charge.
8.8
[1.6] Save parameter set
8.8.1
Parameter sets overview
You can use this menu to save your latest settings in a parameter set. You can use it to
restore any unintentional changes to the parameter assignment. SIPROCESS GA700
devices use two types of parameter sets:
● Factory data: This parameter set is the factory default and is used to restore the original
parameter assignment.
For this reason, factory data can only be loaded.
● User data: This parameter set maps a current, application-specific parameter assignment.
You can save and reload user data at any time.
8.8.2
Save parameter set
Procedure
1. "1 Quick Start" > "6. Save parameter set" main menu
Use this function to save the current parameter assignment in the device. The data is
saved as "user data".
2. Start save process: "Save parameter set" parameter field.
– Authorize access with Standard PIN, if necessary.
– Follow the instructions in the displayed window.
Quick Start
Compact Operating Instructions, 06/2017, A5E31805656-05
Commissioning: Sequence
8.8 [1.6] Save parameter set
105