Siemens SIPROCESS GA700 Quick Start Manual page 25

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Combination operation with ULTRAMAT 7
Table 3- 2
Operating mode
Pressure operation
Suction operation
CALOMAT 7
Table 3- 3
Operating mode
Pressure opera-
tion
Suction operation
Quick Start
Compact Operating Instructions, 06/2017, A5E31805656-05
Mounting conditions for the sample gas restrictor: ULTRAMAT 7
Sample gas restrictor/clamping screw
Combined operation
OXYMAT 7/ULTRAMAT 7
1. OXYMAT 7 must always be
purged first.
2. Install sample gas restrictor in
front of the sample gas inlet of
the OXYMAT 7.
1. ULTRAMAT 7 must always be
purged first.
2. Install sample gas restrictor in
the sample gas outlet of the
OXYMAT 7.
3. Provide sample gas inlet with
clamping screw.
Mounting conditions for the sample gas restrictor: CALOMAT 7
Sample gas restrictor/clamping screw
Operation with one CALOMAT 7
Pre-assembled clamping screws.
No retrofitting or dismantling.
Combination operation
3.3 Gas connections
Combined operation
ULTRAMAT 7/ULTRAMAT 7
No measures
No measures
Combined operation with
ULTRAMAT 7/OXYMAT 7/CALOMAT 7
No restrictions.
When operating in combination with
ULTRAMAT 7, we recommend purging
the CALOMAT 7 as the first module
25