Test Functions and Calibration
Dark Current Test

Test Failed

Probable cause
1
2
3
162
Figure 50
Dark Current Test – Signals
Defective slit assembly (stray light).
Defective detector main board.
Defective PDA/optical unit.
Suggested actions
Run the
"Slit Test (G4212A)"
of the
"Self-Test"
on page 138).
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Agilent 1200 Infinity Series DAD User Manual
on page 151 (part