ResMed AirStart 10 APAP User Manual page 18

Hide thumbs Also See for AirStart 10 APAP:
Table of Contents

Advertisement

Available languages

Available languages

Measurement system uncertainties
In accordance with ISO 80601-2-70:2015 the measurement uncertainty of the manufacturer's test equipment is:
For measures of flow
For measures of volume (< 100 mL)
For measures of volume (≥ 100 mL)
For measures of static pressure
For measures of dynamic pressure
For measures of time
Note: ISO 80601-2-70:2015 stated accuracies and test results provided in this manual for these items already include the
relevant measurement uncertainty from the table above.
Pressure accuracy
Maximum static pressure variation at 10 cm H
Without humidification
With humidification
Maximum dynamic pressure variation according to ISO 80601-2-70:2015
Device without humidification and Standard air tubing / Device with humidification and Standard air tubing
Pressure [cm H
O (hPa)]
2
4
8
12
16
20
Device without humidification and SlimLine air tubing / Device with humidification and SlimLine air tubing
Pressure [cm H
O (hPa)]
2
4
8
12
16
20
Flow (maximum) at set pressures
The following are measured accordingly to ISO 80601-2-70:2015 at the end of the specified air tubing:
Pressure
cm H
O (hPa)
2
4
8
12
16
20
± 1.5 L/min or ± 2.7% of reading (whichever is greater)
± 5 mL or 6% of reading (whichever is greater)
± 20 mL or 3% of reading (whichever is greater)
± 0.15 cm H
± 0.27 cm H
± 10 ms
O (10 hPa) according to ISO 80601-2-70:2015
2
Standard air tubing
± 0.5 cm H
O (± 0.5 hPa)
2
± 0.5 cm H
O (± 0.5 hPa)
2
10 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
10 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
AirStart 10 and Standard
L/min
180
168
157
144
131
O (hPa)
2
O (hPa)
2
SlimLine air tubing
± 0.5 cm H
± 0.5 cm H
15 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
15 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
AirStart 10, humidification
and Standard
L/min
143
135
136
134
123
O (± 0.5 hPa)
2
O (± 0.5 hPa)
2
20 BPM
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
20 BPM
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
AirStart 10 and SlimLine
L/min
162
151
140
128
117
English
17

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Airstart 10 cpap

Table of Contents