ADJUSTMENTS AND REPLACEMENTS
Warning
Dangerous voltage.
[1] De-energize the PROCESSOR.
[2] Remove:
• COVER protecting the BOARDS
• DRIVE CHAIN COVER
• BRACKET with the 2 BOARDS
[3] Remove the desired BOARD.
[4] When you install the new BOARDS and the BRACKET, route the wires from CONNECTORS 7001 and 7002
through the notch in the BRACKET.
Removing the LEVEL SENSOR HOUSING and the LEVEL SENSOR
PROBES
SCREW
PROBE SCREW
ESD
Possible damage from electrostatic discharge.
Corrosion deposits on the LEVEL SENSOR parts may prevent the level sense circuit from operating correctly. Clean
or replace parts as necessary.
[1] See the illustrations to remove the desired part.
[2] When you install the new part, check that the PROBE WIRE is correctly attached to the SCREW and WASHER.
Install the GASKET SIDE of the WASHER against the TANK. See the figure.
32
REPLENISHER
HOSE
MANDREL
ASSEMBLY
LEVEL
SENSOR
HOUSING
CLIP
LEVEL
SENSOR
PROBE
H104_0254ACA
H104_0254AA
GASKET SIDE
WASHER
SCREW
PROBE
WIRE
TANK
H127_0153ACA
H127_0153AA
28JAN98 – 7C8285