Electrical Specifications - GE Venue 50 Basic Service Manual

Hide thumbs Also See for Venue 50:
Table of Contents

Advertisement

System Setup

Electrical specifications

WARNING
Verification of the system's voltage setting
Electrical specifications for Venue 50
Part
Number
Description
5448430
Venue 50 Console
3-30
Connecting a Venue 50 to the wrong voltage level will most
likely destroy it.
Verify that the mains voltage specified for the Venue 50 is
available on-site.
The voltage setting for the unit is found on the rating plate of the
system.
In the table below, the electrical specifications for Venue 50
includes on board peripherals.
Table 3-8:
Electrical specifications for Venue 50
Voltage
100-240 VAC
Figure 3-19. Rating Plate Label
Tolerances
consumption
+ 10%
1.6A max.
Venue 50 – Basic Service Manual
Power
Frequency
50/60Hz
5447566-100 English Rev. 7