Required facility needs
Desirable features
Venue 50 – Basic Service Manual
5447566-100 English Rev. 7
•
Dedicated single branch power outlet of adequate
amperage meeting all local and national codes which is
located less than 2.5 m (8 ft.) from the unit's proposed
location
•
Door opening is at least 76 cm (30 in) wide
•
Proposed location for unit is at least 0.2m (0.67 ft.) from the
wall for cooling
•
Power outlet and place for any external peripheral are within
2 m (6.5 ft.) of each other with peripheral within 1 m of the
unit to connect cables.
•
Power outlets for other medical equipment and gel warmer
•
Power outlets for test equipment and modem within 1 m (3.2
ft.) of unit
•
Clean and protected space to store transducers (in their
cases or on a rack)
•
Material to safely clean probes (done with a plastic
container, never metal)
See 'Electrical requirements' on page 2-6 for more information.
•
Circuit breaker for dedicated power outlet is easily
accessible
•
Sink with hot and cold water
•
Receptacle for bio–hazardous waste, like used probe
sheaths
•
Emergency oxygen supply
•
Storage for linens and equipment
•
Nearby waiting room, lavatory, and dressing room
•
Dual level lighting (bright and dim)
•
Lockable cabinet ordered by GE for its software and
proprietary manuals