AMI Deltacon Power
Default Values
Relay 1/2
80
Sample Temp: (Temp. 1 and Temp. 2)
Alarm High:.............................................................................. 160 °C
Alarm Low:................................................................................... 0 °C
Sample Flow:
Flow Alarm.....................................................................................yes
Alarm high: ................................................................................ 16 l/h
Alarm low:.................................................................................... 5 l/h
Case temp. high: ....................................................................... 65 °C
Case temp. low:........................................................................... 0 °C
Function:.............................................................................limit upper
Parameter:.........................................................................Cond 1(sc)
Setpoint: ................................................................................ 1000 µS
Hysteresis:................................................................................. 10 µS
Delay: ...........................................................................................30 s
If Function = Control upw. or dnw:
Parameter:.........................................................................Cond 1(sc)
............................................................. Settings: Actuator: Frequency
.................................................... Settings: Pulse Frequency:120/min
.................................Settings: Control Parameters: Setpoint:1000 µS
...................................... Settings: Control Parameters: P-band:10 µS
..................................... Settings: Control Parameters: Reset time:0 s
..............................Settings: Control Parameters: Derivative Time:0 s
......................... Settings: Control Parameters: Control Timeout:0 min
Settings: Actuator ............................................................. Time prop.
...................................................................................Cycle time: 60 s
............................................................................Response time: 10 s
Settings: Actuator .............................................................Motor valve
..................................................................................... Run time: 60 s
.................................................................................Neutral zone: 5%
If Function = Timer:
Mode: ......................................................................................Interval
.......................................................................................Interval:1 min
Mode: .......................................................................................... daily
........................................................................ Starting time: 00.00.00
Mode: ...................................................................................... weekly
........................................................................ Starting time: 00.00.00
.......................................................... Calendar; Monday to Sundayoff
Run time: ......................................................................................10 s
Delay: .............................................................................................5 s
Signal output:...............................................................................cont.
Output/Control: ............................................................................cont.
A-96.250.451 / 190121