Swann AMI Deltacon Power Operator's Manual page 63

Monitoring system
Table of Contents

Advertisement

AMI Deltacon Power
Program List and Explanations
5.1.2.2.2
5.1.2.2.3
5.1.2.2.5
5.1.2.2.5.1
5.1.3
5.1.3.1
5.2 Signal Outputs
5.2.1
5.2.1.1
5.2.1.2
5.2.1.3
A-96.250.451 / 190121
Temp. Corr: Enter the temperature correction printed on the sensor
label.
Cable length: Enter the cable length. Set the cable length to 0.0 m if
the sensors are installed in the flow cell on the AMI monitor.
Temp. comp:
Comp.: Available compensation models:
 Strong acids
Flow
Sensor: Choose between <Yes> od <No>. <Yes> is set by default. If
Sensor is set to <No>, the flow monitoring is deactivated and the sig-
nal <Sample flow> on the Signal Outputs and Relay 1 and 2 is not
available.
Note: The navigation in the menu <Signal Output 1> and <Signal
Output 2> is equal. For reason of simplicity only the menu
numbers of Signal Output 1 are used in the following.
Signal Output 1: Assign process value, the current loop range and a
function to each signal output.
Parameter: Assign one of the process values to the signal output.
Available values:
 Cond 1 (cc)
 Cond 2 (sc)
 Temp. 1
 Temp. 2
 Difference
 Sample flow
 pH
 Ammonia
Current Loop: Select the current range of the signal output.
Make sure the connected device works with the same current range.
Available ranges: 0–20 mA or 4–20 mA
Function: Define if the signal output is used to transmit a process val-
ue or to drive a control unit. Available functions are:
 Linear, bilinear or logarithmic for process values.
See
As process values, p. 64
 Control upwards or control downwards for controllers.
See
As control output, p. 65
63