AMI Deltacon Power
Program List and Explanations
5.3.1.3
5.3.1.3.1
5.3.1.3.2
5.3.1.3.35
5.3.1.4
5.3.1.4.1
5.3.1.4.2
5.3.2 and 5.3.3
A-96.250.451 / 190121
Sample Flow: Define at which sample flow an alarm should be is-
sued.
Flow Alarm: Program if the alarm relay should be activated if there is
a flow alarm. Choose between yes or no. The flow alarm will always
be indicated in the display, pending error list, saved in the message
list and the logger. Available values: Yes or no
Note: Sufficient flow is essential for a correct measurement.
We recommend to program yes.
Alarm High: If the measuring values rises above the programmed
value E009 will be issued.
Range: 9–20 l/h
Alarm Low: If the measuring values falls below the programmed val-
ue E010 will be issued.
Range: 5–8 l/h
Case Temp.
Alarm high: Set the alarm high value for temperature of electronics
housing. If the value rises above the programmed value E013 is is-
sued.
Range: 30–75 °C
Alarm low: Set the alarm low value for temperature of electronics
housing. If the value falls below the programmed value E014 is is-
sued.
Range: -10 to +20 °C
Relay 1 and 2: The contacts can be set as normally open or normal-
ly closed with a jumper. See
The function of relay contacts 1 or 2 is defined by the user.
Note: The navigation in the menu <Relay 1> and <Relay 2> is
equal. For reason of simplicity only the menu numbers of Relay 1
are used in the following.
1 First select the functions as:
– Limit upper /lower
– Control upwards /downwards
– Timer
– Fieldbus
2 Enter the necessary data depending on the selected function.
The same values may also be entered in menu
tacts, p. 60
Relay 1 and 2, p.
26.
4.2 Relay Con-
71