Swann AMI Deltacon Power Operator's Manual page 71

Monitoring system
Table of Contents

Advertisement

AMI Deltacon Power
Program List and Explanations
5.3.1.3
5.3.1.3.1
5.3.1.3.2
5.3.1.3.35
5.3.1.4
5.3.1.4.1
5.3.1.4.2
5.3.2 and 5.3.3
A-96.250.451 / 190121
Sample Flow: Define at which sample flow an alarm should be is-
sued.
Flow Alarm: Program if the alarm relay should be activated if there is
a flow alarm. Choose between yes or no. The flow alarm will always
be indicated in the display, pending error list, saved in the message
list and the logger. Available values: Yes or no
Note: Sufficient flow is essential for a correct measurement.
We recommend to program yes.
Alarm High: If the measuring values rises above the programmed
value E009 will be issued.
Range: 9–20 l/h
Alarm Low: If the measuring values falls below the programmed val-
ue E010 will be issued.
Range: 5–8 l/h
Case Temp.
Alarm high: Set the alarm high value for temperature of electronics
housing. If the value rises above the programmed value E013 is is-
sued.
Range: 30–75 °C
Alarm low: Set the alarm low value for temperature of electronics
housing. If the value falls below the programmed value E014 is is-
sued.
Range: -10 to +20 °C
Relay 1 and 2: The contacts can be set as normally open or normal-
ly closed with a jumper. See
The function of relay contacts 1 or 2 is defined by the user.
Note: The navigation in the menu <Relay 1> and <Relay 2> is
equal. For reason of simplicity only the menu numbers of Relay 1
are used in the following.
1 First select the functions as:
– Limit upper /lower
– Control upwards /downwards
– Timer
– Fieldbus
2 Enter the necessary data depending on the selected function.
The same values may also be entered in menu
tacts, p. 60
Relay 1 and 2, p.
26.
4.2 Relay Con-
71