Swann AMI Deltacon Power Operator's Manual page 70

Monitoring system
Table of Contents

Advertisement

AMI Deltacon Power
Program List and Explanations
5.3.1.1.5
5.3.1.1.5.1
5.3.1.1.5.25
5.3.1.1.5.35
5.3.1.1.5.45
5.3.1.2
5.3.1.2.1
5.3.1.2.1.1
5.3.1.2.1.25
5.3.1.2.2
5.3.1.2.2.1
5.3.1.2.2.25
70
Ammonia (if Calculations = yes)
Alarm High: If the measured value rises above the alarm high value,
the alarm relay is activated and E035, is displayed in the message
list.
Range: 0.00–500 ppm
Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E036 is displayed in the message
list.
Range: 0.00–500 ppm
Hysteresis: Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluctu-
ates around the alarm value.
Range: 0.00–500 ppm
Delay: Duration, the activation of the alarm relay is retarded after the
measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28'800 Sec
Sample Temp.
Temp. 1
Alarm High: If the measured value rises above the alarm high value,
the alarm relay is activated and E007, is displayed in the message
list.
Range: 30–200 °C
Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E008 is displayed in the message
list.
Range: -10 to + 20 °C
Temp. 2
Alarm High: If the measured value rises above the alarm high value,
the alarm relay is activated and E037, is displayed in the message
list.
Range: 30–200 °C
Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E038 is displayed in the message
list.
Range: -10 to + 20 °C
A-96.250.451 / 190121