70
Ammonia (if Calculations = yes)
Alarm High: If the measured value rises above the alarm high value,
the alarm relay is activated and E035, is displayed in the message
list.
Range: 0.00–500 ppm
Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E036 is displayed in the message
list.
Range: 0.00–500 ppm
Hysteresis: Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluctu-
ates around the alarm value.
Range: 0.00–500 ppm
Delay: Duration, the activation of the alarm relay is retarded after the
measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28'800 Sec
Sample Temp.
Temp. 1
Alarm High: If the measured value rises above the alarm high value,
the alarm relay is activated and E007, is displayed in the message
list.
Range: 30–200 °C
Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E008 is displayed in the message
list.
Range: -10 to + 20 °C
Temp. 2
Alarm High: If the measured value rises above the alarm high value,
the alarm relay is activated and E037, is displayed in the message
list.
Range: 30–200 °C
Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E038 is displayed in the message
list.
Range: -10 to + 20 °C