Swann AMI Deltacon Power Operator's Manual page 49

Monitoring system
Table of Contents

Advertisement

AMI Deltacon Power
Troubleshooting
Error
Cond. 1 Alarm high
E001
Cond. 1 Alarm low
E002
Cond. 2 Alarm high
E003
Cond. 2 Alarm low
E004
E007
Temp. 1 low
E008
Sample Flow high
E009
Sample Flow low
E010
Temp. 1 shorted
E011
Temp. 1 disconnected
E012
Case Temp. high
E013
E014
Case Temp. low
A-96.250.451 / 190121
Description
Temp. 1 high
Corrective action
– check process
– check programmed value, see
p. 68
– check process
– check programmed value, see
p. 68
– check process
– check programmed value, see
5.3.1.1.2.1, p. 69
– check process
– check programmed value, see
5.3.1.1.2.25, p. 69
– check process
– check programmed value, see
5.3.1.1.4, p. 69
– check process
– check programmed value, see
5.3.1.1.4, p. 69
– check sample inlet pressure
– check programmed value, see
5.3.1.3.2, p. 71
– check sample inlet pressure
– Check flow regulating valve
– check programmed value, see
5.3.1.3.35, p. 71
– Check wiring of temperature sensor
– Check temperature sensor
– Check wiring of temperature sensor
– Check temperature sensor
– check case/environment temperature
– check programmed value, see
5.3.1.4.1, p. 71
– check case/environment temperature
– check programmed value, see
5.3.1.4.2, p. 71
5.3.1.1,
5.3.1.1,
49