Function = Limit upper/lower:
When the relays are used as upper or lower limit switches, program
the following:
Parameter: select a process value
Setpoint: If the measured value rises above respectively falls below
the set-point, the relay is activated.
Parameter
Range
Cond. 1 (sc)
0–3000 S
Cond.2 (cc)
0–3000 S
Temp. 1
-25 to +270 °C
Temp. 2;
-25 to +270 °C
Difference
0–3000 S
Sample flow
0–20 l/h
pH
0 –14 pH
Ammonia
0 –500 ppm
Hysteresis: within the hysteresis range, the relay does not switch.
This prevents damage of relay contacts when the measured value
fluctuates around the alarm value.
Parameter
Range
Cond. 1 (sc)
0–3000 S
Cond.2 (cc)
0–3000 S
Temp. 1
0–100 °C
Temp. 2;
0–100 °C
Difference
0–3000 S
Sample flow
0–20 l/h
pH
0 –14 pH
Ammonia
0 –500 ppm
Delay: Duration, the activation of the alarm relay is retarded after the
measuring value has risen above/fallen below the programmed
alarm.
Range. 0–600 Sec