Active: Define when the input should be active:
No:
Input is never active.
When closed
Input is active if the input relay is closed
When open:
Input is active if the input relay is open
Signal Outputs: Select the operation mode of the signal outputs
when the relay is active:
Continuous:
Signal outputs continue to issue the measured
value.
Hold:
Signal outputs issue the last valid measured value.
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Off:
Set to 0 or 4 mA respectively. Errors, except fatal
errors, are not issued.
Output/Control: (relay or signal output):
Continuous:
Controller continues normally.
Hold:
Controller continues on the last valid value.
Off:
Controller is switched off.
Fault:
No:
No message is issued in pending error list and the
alarm relay does not close when input is active.
Message E024 is stored in the message list.
Yes
Message E024 is issued and stored in the message
list. The Alarm relay closes when input is active.
Delay: Time which the instrument waits, after the input is deactivat-
ed, before returning to normal operation.
Range: 0–6'000 Sec