Product specifications

Displayed temperature range
Operating ambient
temperature range
Display resolution
Accuracy for patient temperature range
96.8 °F -102.2 °F (36 °C - 39 °C): ± 0.4 °F
outside this range:
Long term storage ranges
Temperature:
Humidity:
Battery life:
This infrared thermometer meets requirements established in ASTM
Standard E 1965-98 (for the thermometer system [thermometer with
lens filter]). Full responsibility for the conformance of the product to the
standard is assumed by Braun GmbH, 61476 Kronberg, Germany.
ASTM laboratory accuracy requirements for the thermometer only (not
including a lens filter) in the display range of 98.8 °F to 102.2 °F
(36 °C to 39 °C) for infrared thermometers is ± 0.4 °F (± 0.2 °C), whereas
for mercury-in-glass and electronic thermometers, the requirement per
ASTM Standards E 667-86 and E 1112-86 is ± 0.2 °F ( ± 0.1 °C).
This appliance conforms to the following standards:
DIN EN 60601-1: 3/96 «Medical electrical equipment» –
Part 1: General requirements for safety
DIN EN 12470-5: 2003 «Clinical thermometers» –
Part 5: Performance of infrared ear thermometers
(with maximum device)
TYPE BF
Attention, consult
EQUIPMENT
ACCOMPANYING
DOCUMENTS
Internally Powered Equipment
Continuous Operation
Not Protected against Ingress of Water
U.S. Patent No. 5,088,834 Other Patents Pending
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC. For
detailed description of EMC requirements please contact the Braun ThermoScan Customer
Service at 1-800-327-7226.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
In the unlikely event you experience any difficulty using the Braun ThermoScan thermometer,
simply call us toll free at 1-800-327-7226.
93.2 °F -108 °F (34 °C - 42.2 °C)
50 °F -104 °F
0.1 °F or °C
Maximum Laboratory Error
± 0.5 °F
- 4 °F to 122 °F ( - 20 °C to 50 °C)
95 % non-condensing
2 years / 1000 measurements
MEDICAL ELECTRICAL EQUIPMENT
CLASSIFIED BY UNDERWRITERS LABORATORIES INC.
WITH RESPECT TO ELECTRIC SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL 2601-1 / CAN CSA C22.2
No.601.1 10HK
(10 °C - 40 °C)
± 0.2 °C