6 Preparing for a run

Performing pre-run checks

Performing pre-run checks
About this topic
Procedure
Related information
128
Before you start a protocol run, you should check the BenchCel Microplate Handler to
ensure optimum operation. Perform the following procedure before each protocol run.
To check that the BenchCel and integrated devices are ready for a run:
1
Ensure the labware racks are filled with the correct labware.
2
Make sure there are no stray microplates in robot-accessible locations, including
platepads or plate stages on integrated devices.
3
Remove any obstacle in the robot's pathways. Consider the path of the robot head
and arms.
4
For the devices that require compressed air, make sure the air pressure meets the
operating requirements.
To check the BenchCel internal air pressure, see
pressure" on page
158.
5
Make sure all devices and accessories are set up correctly and prepared for the run.
For instructions, see the user documentation for the device or accessory.
For information about...
BenchCel air source requirements
Integrated device air source
requirements
Diagnosing air pressure problems
"Adjusting the stacker gripper
See...
"Verifying laboratory requirements" on
page 37
External device user documentation
"Maintenance and troubleshooting" on
page 129
BenchCel Microplate Handler User Guide