Cbrn Scba Or Apr; Fit Testing Procedure; Negative Pressure Leak Test; Quantitative Fit Testing - 3M SCOTT AIR-PAK X3 PRO Operating & Maintenance Instructions

Self-contained breathing apparatus (scba), nfpa-1981 (2018 edition) compliant
Hide thumbs Also See for SCOTT AIR-PAK X3 PRO:
Table of Contents

Advertisement

When using a PortaCount Respirator Fit Tester for quantitative fit testing, TSI recommends that the level of particles
in the ambient air must be between 5000 and 30000 particles/cm
level of ambient particle count required, refer to the PortaCount Respirator Fit Tester user instructions for details
including available Particle Generators to use with the PortaCount Respirator Fit Tester. When using the controlled
negative pressure protocol, model specific adapters must be used. These are provided by the manufacturer of the
controlled negative pressure fit test device.

CBRN SCBA or APR

If you are fit testing for CBRN SCBA Open Circuit mode of operation (minimum fit factor equal to or greater than 500
minimum) or CBRN APR (minimum fit factor equal to or greater than 2000 minimum), appropriate negative pressure
testing equipment must be used. For example:
If you are fit testing for CBRN SCBA Open Circuit or CBRN APR, use a P100 Filter (P/N 052683) and Fit Test
Adapter (P/N 201520-01).
If you are using a facepiece that has been specially modified for a fit testing program such as that described in
NIOSH Procedure No. RB-CET-STP-CBRN-0352, Determination of Laboratory Respirator Protection Level
(LRPL) Quantitative, Medium Flow, Deep Probe, Corn Oil, Fit Factor Performance Test For Chemical, Biological,
Radiological and Nuclear (CBRN) Full Facepiece Respiratory Protection Devices (RPD) Standard Testing
Procedure, install the appropriate filtration element directly into the 40 mm facepiece adapter.

Fit Testing Procedure

Negative Pressure Leak Test

The size and style facepiece must be selected based on the user's measured face size. For the initial fitting of the
facepiece, conduct a negative pressure leak test as described in
procedure carefully. If the selected facepiece does not pass the negative pressure leak test or does not fit securely
without movement in the chin or chin cup area, or if the user experiences discomfort in the chin or throat, try the next
nearest size, larger or smaller. After passing the negative pressure leak test, the facepiece size selected must be
verified by successfully passing a respirator quantitative fit test.
A leak that is detected from a location other than the face-to-facepiece seal may indicate damaged or defective
equipment. Remove the defective equipment from service and tag it for repair by authorized personnel. Repeat the
testing with equipment known to be operating properly.

Quantitative Fit Testing

Test subjects must be in good health at the time of the fit testing. Smoking or eating less than 30 minutes prior to the
test is prohibited. Any and all conditions that might interfere with a good face-to-facepiece seal must be addressed
and corrected before performing the fit testing.
To verify the fit factor of the respirator, testing must incorporate an exercise regimen of normal daily activities. 3M
requires the following set of fit test exercises, which are based on OSHA Standard 29 CFR Part 1910.134 Appendix
A, and ANSI/AIHA/ASSE Z88.10, current edition, with modifications.
Each exercise is to be performed for 60 seconds (except as noted) in a standing position during the test:
Normal breathing
Deep breathing
Turning head side to side
Moving head up and down (look up/look down)
Talking (read "The Rainbow Passage")
60
. If you have difficulty achieving the minimum
3
"Starting the
Respirator" on page 23. Follow the
P/N 595373-01
Rev C
202004

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents