Remote Cables - Agilent Technologies 1290 Infinity User Manual

Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

8
Cable Overview

Remote Cables

Remote Cables
One end of these cables provides a Agilent Technologies APG (Analytical
Products Group) remote connector to be connected to Agilent modules. The
other end depends on the instrument to be connected to.
Agilent Module to 3396A Integrators
p/n 03394-60600
Agilent Module to 3396 Series II / 3395A Integrators
Use the cable Agilent module to 3396A Series I integrators (p/n 03394-60600)
and cut pin #5 on the integrator side. Otherwise the integrator prints START;
not ready.
78
Pin 3394
Pin Agilent
module
9
1 - White
NC
2 - Brown
3
3 - Gray
NC
4 - Blue
NC
5 - Pink
NC
6 - Yellow
5,14
7 - Red
1
8 - Green
NC
9 - Black
13, 15
1290 Infinity Thermostat User Manual
Signal Name
Active
(TTL)
Digital ground
Prepare run
Low
Start
Low
Shut down
Low
Not connected
Power on
High
Ready
High
Stop
Low
Start request
Low
Not connected