Shop for Fluke products online at:
CH2 Level 35 MHz
CH2 Level 37·5 MHz
CH2 Level 40 MHz
CH3 DC offset zero.
CH3 DC offset at + full scale.
CH3 DC offset at − full scale.
CH3 Multiplier zero.
CH3 Multiplier offset.
CH3 Waveform offset.
CH3 Output level at full scale
CH3 20 dB attenuator
CH3 40 dB attenuator
CH3 10 dB attenuator
CH3 Sum offset
CH3 SCM level at full-scale
CH3 AM level at full-scale
CH3 Level 0.1 MHz
CH3 Level 33 MHz
CH3 Level 1 MHz
CH3 Level 2 MHz
CH3 Level 4 MHz
CH3 Level 5 MHz
CH3 Level 10 MHz
CH3 Level 15 MHz
CH3 Level 20 MHz
CH3 Level 25 MHz
CH3 Level 30 MHz
CH3 Level 32·5 MHz
CH3 Level 35 MHz
CH3 Level 37·5 MHz
CH3 Level 40 MHz
CH4 DC offset zero.
CH4 DC offset at + full scale.
CH4 DC offset at − full scale.
CH4 Multiplier zero.
CH4 Multiplier offset.
CH4 Waveform offset.
CH4 Output level at full scale
CH4 20 dB attenuator
CH4 40 dB attenuator
CH4 10 dB attenuator
CH4 Sum offset
CH4 SCM level at full-scale
CH4 AM level at full-scale
CH4 Level 0.1 MHz
CH4 Level 33 MHz
CH4 Level 1 MHz
CH4 Level 2 MHz
CH4 Level 4 MHz
CH4 Level 5 MHz
CH4 Level 10 MHz
CH4 Level 15 MHz
CH4 Level 20 MHz
CH4 Level 25 MHz
CH4 Level 30 MHz
MyFlukeStore
www.
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for 0 V ±5 mV
Adjust for +10 V ±10 mV
Check for –10 V ±3 %
Adjust for minimum volts ac
Adjust for 0 V ±5 mV
Adjust for 0 V ±5 mV
Adjust for 10 V ±10 mV
Adjust for 1 V ±1 mV
Adjust for 0·1 V ±·1 mV
Adjust for 2·236 V ac ±10 mV
Adjust for 0 V ±5 mV
Adjust for 5 V ±5 mV
Adjust for 10 V ±10 mV
Note reading
Check reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for 0 V ±5 mV
Adjust for +10 V ±10 mV
Check for –10 V ±3 %
Adjust for minimum volts ac
Adjust for 0 V ±5 mV
Adjust for 0 V ±5 mV
Adjust for 10 V ±10 mV
Adjust for 1 V ±1 mV
Adjust for 0·1 V ±·1 mV
Adjust for 2·236 V ac ±10 mV
Adjust for 0 V ±5 mV
Adjust for 5 V ±5 mV
Adjust for 10 V ±10 mV
Note reading
Check reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
1.888.610.7664
.com
16
Calibration
Calibration Routine
16-5