The full procedure (for the 4-channel instrument) is as follows:
CAL 03
CAL 04
CAL 05
CAL 06
CAL 07
CAL 08
CAL 09
CAL 10
CAL 11
CAL 12
CAL 13
CAL 14
CAL 15
CAL 16
CAL 17
CAL 18
CAL 19
CAL 20
CAL 21
CAL 22
CAL 23
CAL 24
CAL 25
CAL 26
CAL 27
CAL 28
CAL 29
CAL 30
CAL 31
CAL 32
CAL 33
CAL 34
CAL 35
CAL 36
CAL 37
CAL 38
CAL 39
CAL 40
CAL 41
CAL 42
CAL 43
CAL 44
CAL 45
CAL 46
CAL 47
CAL 48
CAL 49
CAL 50
CAL 51
CAL 52
CAL 53
CAL 54
CAL 55
16-4
Shop for Fluke products online at:
CH1 DC offset zero.
CH1 DC offset at + full scale.
CH1 DC offset at − full scale.
CH1 Multiplier zero.
CH1 Multiplier offset.
CH1 Waveform offset.
CH1 Output level at full scale
CH1 20 dB attenuator
CH1 40 dB attenuator
CH1 10 dB attenuator
CH1 Not used.
CH1 Not used.
CH1 Not used.
CH1 Level 0.1 MHz
CH1 Level 33 MHz
CH1 Level 1 MHz
CH1 Level 2 MHz
CH1 Level 4 MHz
CH1 Level 5 MHz
CH1 Level 10 MHz
CH1 Level 15 MHz
CH1 Level 20 MHz
CH1 Level 25 MHz
CH1 Level 30 MHz
CH1 Level 32·5 MHz
CH1 Level 35 MHz
CH1 Level 37·5 MHz
CH1 Level 40 MHz
CH2 DC offset zero.
CH2 DC offset at + full scale.
CH2 DC offset at − full scale.
CH2 Multiplier zero.
CH2 Multiplier offset.
CH2 Waveform offset.
CH2 Output level at full scale
CH2 20 dB attenuator
CH2 40 dB attenuator
CH2 10 dB attenuator
CH2 Sum offset
CH2 SCM level at full-scale
CH2 AM level at full-scale
CH2 Level 0.1 MHz
CH2 Level 33 MHz
CH2 Level 1 MHz
CH2 Level 2 MHz
CH2 Level 4 MHz
CH2 Level 5 MHz
CH2 Level 10 MHz
CH2 Level 15 MHz
CH2 Level 20 MHz
CH2 Level 25 MHz
CH2 Level 30 MHz
CH2 Level 32·5 MHz
MyFlukeStore
www.
Adjust for 0 V ±5 mV
Adjust for +10 V ±10 mV
Check for –10 V ±3 %
Adjust for minimum volts ac
Adjust for 0 V ±5 mV
Adjust for 0 V ±5 mV
Adjust for 10 V ±10 mV
Adjust for 1 V ±1 mV
Adjust for 0·1 V ±·1 mV
Adjust for 2·236 V ac ±10 mV
Note reading
Check reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for 0 V ±5 mV
Adjust for +10 V ±10 mV
Check for –10 V ±3 %
Adjust for minimum volts ac
Adjust for 0 V ±5 mV
Adjust for 0 V ±5 mV
Adjust for 10 V ±10 mV
Adjust for 1 V ±1 mV
Adjust for 0·1 V ±·1 mV
Adjust for 2·236 V ac ±10 mV
Adjust for 0 V ±5 mV
Adjust for 5 V ±5 mV
Adjust for 10 V ±10 mV
Note reading
Check reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading
Adjust for same reading