Specifications - Oxygen; Specifications - Free Chlorine; Specifications - Total Chlorine; Specifications - Ozone - Emerson Rosemount Analytical HART 5081-A-HT Instruction Manual

Two-wire chlorine, dissolved oxygen, and ozone transmitter
Table of Contents

Advertisement

MODEL 5081-A
Explosion-Proof:
Class I, Div. 1, Groups B-D
Class II, Div. 1, Groups E-G
Class III, Div. 1
Class I, Groups B-D
Class II, Groups E-G
Class III
Tamb = 65°C max
1.3 SPECIFICATIONS
Measurement Range: 0-99 ppm (mg/L), 0-200%
saturation
Resolution: 0.01 ppm, 0.1 ppb for 499A TrDO sen-
sor
Temperature correction for membrane perme-
ability: automatic between 0 and 50°C (can be
disabled)
Calibration: air calibration (user must enter baro-
metric pressure) or calibration against a stan-
dard instrument
RECOMMENDED SENSORS — OxyGEN:
Model 499A DO-54 for ppm level
Model 499A TrDO-54 for ppb level
Hx438 and Gx448 steam-sterilizable oxygen sensors
1.4 SPECIFICATIONS
Measurement Range: 0-20 ppm (mg/L) as Cl
Resolution: 0.001 ppm (Autoranges at 0.999 to
1.00 and 9.99 to 10.0)
Temperature correction for membrane perme-
ability: automatic between 0 and 50°C (can be
disabled)
pH Correction: Automatic between pH 6.0 and
9.5. Manual pH correction is also available.
Calibration: against grab sample analyzed using
portable test kit.
RECOMMENDED SENSOR — FREE CHLORINE:
Model 499A CL-01-54
SPECIFICATIONS — pH
Application: pH measurement available with free
chlorine only
Measurement Range: 0-14 pH
Resolution: 0.01 pH
Sensor Diagnostics: Glass impedance (for broken
or aging electrode) and reference offset.
Reference impedance (for fouled reference
junction) is not available.
OxyGEN
FREE CHLORINE
2
Repeatability: ±0.01 pH at 25°C
RECOMMENDED SENSOR — pH
Use Model 399-09-62, 399-14, or 399VP-09.
See pH sensor product data sheet for complete order-
ing information.
1.5 SPECIFICATIONS
Measurement Range: 0-20 ppm (mg/L) as Cl
Resolution: 0.001 ppm (Autoranges at 0.999 to
1.00 and 9.99 to 10.0)
Temperature correction for membrane perme-
ability: automatic between 5 and 35°C (can be
disabled)
Calibration: against grab sample analyzed using
portable test kit.
RECOMMENDED SENSOR — TOTAL CHLORINE:
Model 499A CL-02-54 (must be used with SCS 921)
1.6 SPECIFICATIONS
Measurement Range: 0-10 ppm (mg/L)
Resolution: 0.001 ppm (Autoranges at 0.999 to
1.00 and 9.99 to 10.0)
Temperature correction for membrane perme-
ability: automatic between 5 and 35°C (can be
disabled)
Calibration: against grab sample analyzed using
portable test kit.
RECOMMENDED SENSOR — OzONE:
Model 499A OZ-54
SPECIFICATIONS — PERCENT
1.7
OxyGEN IN GAS
Measurement Range: 0-25% Oxygen
Resolution: 0.1% - TBD
Calibration: air calibration (automatic meas
urement of barometric pressure with internal
pressure sensor) - TBD
Sample Pressure: 0 to 50 PSIG
Sample Temperature: 32 to 110°F
RECOMMENDED SENSOR — PERCENT
OxyGEN IN GAS
Model 4000 Percent Oxygen Sensor
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
:
TOTAL CHLORINE
2
OzONE
3