Factory Default; Hart - Emerson Rosemount Analytical HART 5081-A-HT Instruction Manual

Two-wire chlorine, dissolved oxygen, and ozone transmitter
Table of Contents

Advertisement

MODEL 5081-A

7.6 FACTORy DEFAULT

7.6.1 Purpose
This section describes how to erase ALL user-defined configuration settings and return the transmitter to factory default
settings.
7.6.2 Procedure
PROGRAM
dISPLAY
EXIT
NEXT
ENTER
PROGRAM
tYPE
02
EXIT
ENTER

7.7 HART

For more information, see Section 6.3.2.
1. Press PROG on the remote controller.
2. Press NEXT until the dEFAULt appears in the display. Press ENTER.
3. Use é or ê to toggle between nO and yES. With yES showing, press ENTER to return
to factory default settings.
SECTION 7.0
PROGRAMMING
41