Procedure - Emerson Rosemount Analytical HART 5081-A-HT Instruction Manual

Two-wire chlorine, dissolved oxygen, and ozone transmitter
Table of Contents

Advertisement

MODEL 5081-A

8.2. PROCEDURE

1. Place the sensor and a calibrated reference thermometer in a container of water at ambient temperature. Be
sure the temperature element in the sensor is completely submerged by keeping the sensor tip at least three
inches below the water level. Stir continuously. Allow at least 20 minutes for the standard thermometer, sen-
sor, and water to reach constant temperature.
CALIBRATE
tEMP AdJ
EXIT
NEXT
ENTER
CALIBRATE
tEMP
025 . 0
EXIT
ENTER
50
2. Press CAL on the remote controller.
3. Press NEXT until the tEMP AdJ submenu appears. Press Enter.
4. The tEMP prompt appears. Use the arrow keys to change the display to
match the temperature measured using the standard thermometer.
Press ENTER to save.
5. The tEMP AdJ sub-menu appears. Press RESET to return to the main
display.
SECTION 8.0
CALIBRATION — TEMPERATURE