MODEL 5081-A

15.1 GENERAL

The 5081-A transmitter can display diagnostic information that is useful in troubleshooting. The diagnostics avail-
able depend on the measurement being made. To read diagnostic information, go to the main display and press
DIAG on the infrared remote controller. Press NEXT until the mnemonic for the desired information appears. Refer
to the appropriate section below for more information.

15.2 DIAGNOSTIC MESSAGES FOR DISSOLvED OxyGEN

TyPE O2
Transmitter is measuring oxygen. Press NEXT to view diagnostics.
SEnSor Cur
Press ENTER to display raw current from sensor (note units).
SEnSitvty
Press ENTER to display sensitivity. Sensitivity is calculated during calibration. It is the
measured current divided by concentration.
O CurrEnt
Press ENTER to display the zero current measured during calibration (note units).
bAr PreSS
Press ENTER to display the barometric pressure used by the transmitter during air calibration.
5081-A-Ht
This is the model number. Press ENTER to display the software revision (SFtr) level. Press
NEXT to show the hardware revision (HArdr) level.
FAULtS
Press ENTER to scroll through existing fault messages.

15.3 DIAGNOSTIC MESSAGES FOR OzONE AND TOTAL CHLORINE

TyPE O3 or tCL
Transmitter is measuring ozone (or total chlorine). Press NEXT to view diagnostics.
SEnSor Cur
Press ENTER to display raw current from sensor (note units).
SEnSitvty
Press ENTER to display sensitivity. Sensitivity is calculated during calibration. It is the
measured current divided by concentration.
O CurrEnt
Press ENTER to display the zero current measured during calibration (note units).
5081-A-Ht
This is the model number. Press ENTER to display the software revision (SFtr) level.
Press NEXT to show the hardware revision (HArdr) level.
FAULtS
Press ENTER to scroll through existing fault messages.