MODEL 5081-A

5.2 INFRARED REMOTE CONTROLLER (IRC) - KEy FUNCTIONS

The infrared remote controller is used to calibrate and program the transmitter and to display diagnostic mes-
sages. See Figure 5-3 for a description of the function of the keys.
Hold the IRC within 6 feet of the transmitter, and not more than 15 degrees from the center of the display window.
RESET - Press RESET to end the cur-
rent operation and return to the main
display. Changes will NOT be saved.
RESET does NOT return the transmit-
ter to factory default settings.
ARROW KEyS - Use é and ê keys to
increase or decrease a number or to
scroll through items in a list. Use the ç
keys to move the cursor across a
or è
number. A flashing word or numeral
shows the position of the cursor.
CAL - Press CAL to access the calibra-
tion menu.
PROG - Press PROG to access the pro-
gram menu.
DIAG - Press DIAG to read diagnostic
messages.
24
FIGURE 5-3. Infrared Remote Controller and label.
DISPLAy AND OPERATION
HOLD
- Press HOLD to access the
prompts used for turning on or off the
hold function.
ENTER - Press ENTER to move from a
submenu to the first prompt under the
submenu. Pressing ENTER also stores
changes in memory and advances to the
next prompt.
NExT - Press NEXT to advance to the
next submenu or to leave a message
screen.
ExIT - Press EXIT to end the current
operation. Changes are NOT saved.
SECTION 5.0