Diagnostic Messages For Free Chlorined - Emerson Rosemount Analytical HART 5081-A-HT Instruction Manual

Two-wire chlorine, dissolved oxygen, and ozone transmitter
Table of Contents

Advertisement

MODEL 5081-A
15.4 DIAGNOSTIC MESSAGES FOR FREE CHLORINE
TyPE FCL
Transmitter is measuring free chlorine. Press NEXT to view diagnostics.
SEnSor Cur
Press ENTER to display raw current from sensor (note units).
SEnSitvty
Press ENTER to display sensitivity. Sensitivity is calculated during calibration. It is the measured
current divided by concentration.
O CurrEnt
Press ENTER to display the zero current measured during calibration (note units).
PH
Press ENTER to view pH diagnostics. Press NEXT to skip pH diagnostics.
InPut
Current pH sensor input voltage in millivolts.
SLOPE
Sensor slope in millivolts per unit pH. Slope is calculated during buffer calibration. See Figure 13.1.
OFFSt
Sensor voltage in millivolts in pH 7 buffer.
GIMP
Glass impedance in MW.
5081-A-Ht
This is the model number. Press ENTER to display the software revision (SFtr) level. Press
NEXT to show the hardware revision (HArdr) level.
FAULtS
Press ENTER to scroll through existing fault messages.
SECTION 15.0
DIAGNOSTICS
75