USER CODE SETUP

Submenus
FLOW DIRECTION
VOLUME FLOW
SOUND VELOCITY
8. Commissioning
The user code can be changed in this menu.
The code is factory-set at 1000. If the user code is lost, the factory setting can be re-established
as follows: Switch off the supply voltage, press the TOP-UP key
Release the key after ROM and RAM tests are completed. The user code is now reset to 1000.
If the signal converter is left in SIGNAL CONVERTER SETUP for more than 10 minutes, the
converter automatically reverts to the OPERATOR MENU.
The figure on previous page gives a view of the menu structure. Below is a more detailed description
of the submenus. Use the menu block diagram on previous page to step through to required
submenu. (See also detailed diagram blocks).
Information regarding the SERVICE MENU is located in the section "Trouble shooting".
The flow direction is defined for volume and mass flow simultaneously.
Choose measuring unit and the maximum volume flow.
For SONO 3000 not fitted with a SENSORPROM
formatted. It is done as follows:
a: Sensor ≤ DN 2000:
In the menu "MAX RANGE" the value 10 x "VOL FLOW MAX" should be entered.
b: Sensor > DN 2000:
In the menu "MAX RANGE" the value 5 x "VOL FLOW MAX" should be entered
(Max. 540.000 m
3
/h).
The measuring unit for sound velocity is meters per second. The measuring range is set by entering
the minimum and the maximum sound velocity.
Error relay and error indication are activated if the sound speed is outside the limits stated.
SFIDK.PS.029.V3.02 - 521H0553
VOLUME
FLOW
VOLUME FLOW
MAX RANGE
3
UNIT: m
/s
UNIT: m
3
/min
UNIT: m
3
/h
UNIT: l/s
UNIT: l/min
UNIT: l/h
UNIT: ml/s
UNIT: UGPS
UNIT: UGPM
UNIT: UGPH
UNIT: MUGPO
UNIT: Ml/d
®
memory unit the display indication must be
SOUND
VELOCITY
SOUND
SOUND
VELOCITY
VELOCITY
MIN.
MAX.
and switch on the supply voltage.
VOL. FLOW
MAX.
39