Installation In Concrete Pipe - Siemens SITRANS F US SONOFLO Handbook

Ultrasonic flowmeter type sonokit 1-track dn 100 - dn 4000
Hide thumbs Also See for SITRANS F US SONOFLO:
Table of Contents

Advertisement

® ® ® ® ®
SITRANS F US SONOFLO

Installation in concrete pipe

18
3. Construction of the sensor
Step 5
Step 6
SFIDK.PS.029.V3.02 - 521H0553
Mark a point on the transducer holders marking
line at a distance c.
c is determined by measuring the wall thickness
t and the mounting plate m.
Mount the transducer holders so that they are
flush with the inner surface of the pipe.
c is calculated using the following formula:
c = (t + m) x 1.15
The correctly positioned transducer holders are
tack-welded in at least 3 places with mandrels
inserted.
Remove the mandrel and fully weld the trans-
ducer holders. Make sure that the transducer
holders are bent as little as possible by the
heat generated during welding.
Check whether the mandrel can still pass through
the transducer holders. Correct the alignment of
the transducer holders if necessary. Use a 1½"
protection nut to protect the transducer holders'
thread and a hammer.
Step 1
Measuring and marking as for steel pipes using
the drawing paper supplied with the mounting
kit or directly on the pipe, see above.
Once points P
and P
are marked, the holes
1
2
can be made.
Step 2
Drill 2 holes, ø 46 mm, into the wall of the pipe
at an angle of 60°, advantageously using the
drill unit shown on the left. The unit is not part of
the mounting kit, but can be rented from
companies specializing in construction equip-
ment.
The unit contains a core drill bit, an ordinary
handheld drill and a device for fastening the unit
to the pipe wall.
The unit is fastened to the pipe wall by means of
vacuum or clamping band.