4. Measuring the sensor
To allow theoretical calibration by the SONO 3000 signal converter the sensor must be measured.
The following sensor data must be known:
⇒ θ = Angle between sound track and longitudinal axis of the pipe
⇒ L = Distance between transducer windows
⇒ Di = Inner diameter of the pipe
The inaccuracies with which θ, L and Di are measured directly influence the measuring accuracy
of the finished system during theoretical calibration. Be sure to choose measuring equipment with
a suitable measuring accuracy.
θ
DN 400-3000
∆L
DN 400-1000
DN 1000-2000
DN 2000-3000
∆Di
DN 400-1000
DN 1000-2000
DN 2000-3000
Step 1
Step 2
SFIDK.PS.029.V3.02 - 521H0553
The table shows the required measuring
0.1°
accuracy:
0.8 mm
2.0 mm
4.0 mm
0.8 mm
2.0 mm
4.0 mm
Measuring the angle ( θ )
Measure the angle at each transducer holder
several times and calculate the average of all
measurements at each transducer holder.
Enter angle θ of each transducer holder into the
measuring report. Calculate the mean angle of
each sound track and enter the value into the
measuring report.
Distance between transducers (L)
Measure L for each sound track and enter the
value into the measuring report.
L may be measured by inserting the mandrel or
a similar instrument into the transducer holders,
which makes it possible to measure the distance
between the thread ends of both transducer
holders, see drawing.