Installation Of The Signal Converter; Installation Of The Transducer Type Sono 3200 - Siemens SITRANS F US SONOFLO Handbook

Ultrasonic flowmeter type sonokit 1-track dn 100 - dn 4000
Hide thumbs Also See for SITRANS F US SONOFLO:
Table of Contents

Advertisement

® ® ® ® ®
SITRANS F US SONOFLO
Installation of the trans-
ducer type SONO 3200
24

6. Installation of the signal converter

The transducer type SONO 3200 is used together with the sensor type SONO 3100 and the
mounting kit type SONOKIT.
O-ring type:
The transducer has a screwed union connection for fitting in a counterflange welded onto the pipe.
The union and the pipe are sealed with an O-ring.
The transducer holders are welded directly onto steel pipes or cast into the walls of concrete pipes
and flush with the internal walls.
The terminal box is fastened to the transducer holder by means of a union. Stress on the connection
wires must be avoided when passing them through the terminal box. The wire without a spade
terminal is connected to terminal 2. The wire with a spade terminal is connected to the transducer
by means of screw 3. The inner wire is connected to terminal 1 and the transducer cable and screen
are fixed to cable clip 4.
Note:
When disassembling the terminal box the wires must be loosened from the connection board. It
is not necessary to remove the transducer cable.
Support the terminal box while loosening the union. Then remove the terminal box from the
transducer holder.
Turning the terminal box can damage the wires to the crystals.
SFIDK.PS.029.V3.02 - 521H0553