Patient Leakage Current Tests; Patient Leakage Current To Ground - GE MAC 600 SPY Service Manual

Table of Contents

Advertisement

Patient Leakage Current Tests

Patient Leakage Current to Ground

4-26
Maintenance: Check out Procedure
N.C.
500 uA (IEC)
150 uA (UL)
These tests measure leakage current of the device's floating inputs
(patient connections ground).
This test measures leakage current from the floating input (patient
connection) to ground during normal condition (N.C.) and single fault
condition (S.F.C). In all cases, the leakage current is measured from the
floating input to ground: The device has to be turned on and off, and
connected to your safety testing instrument.
During N.C., measurements should be taken under the following
conditions (refer to
"Electrical Safety Tests"
Device in ON state
Polarity switch NORM and RVS
GND switch GND closed
S1 (neutral) closed
During S.F.C., measurements should be taken under the following
conditions (refer to
"Electrical Safety Tests"
Device in ON state
Polarity switch NORM and RVS
GND switch closed, S1 (neutral) open
GND switch open, S1 (neutral) closed
TM
MAC
600
2047426-002
S.F.C
1,000 uA (IEC & UL)
on page 4-24):
on page 4-24):
Revision G