Required Tools And Manuals; Pwa Connector Details - GE MAC 600 SPY Service Manual

Table of Contents

Advertisement

Required Tools and Manuals

PWA Connector Details

4-4
Maintenance: Preparing System for FRU Replacement
Sl no:
The legends for the PWA Connectors (J1, J2 etc) for the new PWA FRU
2066447-002 are different as compared with the old PWA FRU 2047330-
001. Refer the following table for the old and new connector legends:
Connector Description
Key sheet
Display
Printer
SD Card
Power In
RS232
JATAG
USB
Modem
TM
MAC
600
2047426-002
Table 4-1. Required tools and manuals
1
Phillips screwdriver
2
Anti-static wrist strap
3
Service manual as needed for reference
4
Safety tester for measurements according to 60601-1
5
Connecting wires and clips required for electrical
safety tests (optionl)
6
Multi-meter
7
Multi Lead ECG simulator
8
MAC 600 Operator's Manual for reference
9
Nose Plier
10
Recommended SD card
Table 4-2. PWA Connectors
OLD PWA FRU
PN: 2047330-001
J4
J5
J10
J1
W1
J2
J3
J9
J12
Description
NEW PWA FRU
PN:2066447-002
J7
J6
J5
J3
W1
J2
J8
J1
J9
Revision G