Earth Cable Routing - GE MAC 600 SPY Service Manual

Table of Contents

Advertisement

Earth Cable Routing

4-12
Maintenance: Preparing System for FRU Replacement
Verify the earth cable routing is as shown in the following figure. It
should connect screw A and screw B. This is the Class II Earth (ground)
cable routing. To change the cable routing to comply with Class I
requirements, perform the following procedure.
1. Open the device and printer door as described in
Device"
on page 4-8.
2. Remove the screws from location A and B.
3. Remove the earth cable (2033817-001).
4. Using the cable and screws from steps 2 and 3, place the cable on
location C and D. The other end D can be connected to either of the
screw points on the printer base. Ensure the cable does not touch any
components on the PCB board and secure it in place with the screws.
TM
MAC
600
2047426-002
Table 4-7. Routing the Earth Cable
Legend
Description
1
Assembly of the earthing cable with main
PCB assembly
2
Earthing cable assembly location
A and B
Class II Earth cable routing
C and D
Class I Earth cable routing
"Opening the