Electrical Safety Tests

Recommended Test Equipment

Recommended Test Schedule

Ground Resistance Test

Leakage Current Tests

4-24
Maintenance: Check out Procedure
The suggested safety analysis tests comply to the international standard
IEC 60601-1. The tests should be performed with safety testing
instruments with measuring circuits calibrated according to IEC 60601-
1. To ensure personal safety, refer to the "testing instrument's user
documentation".
Perform the tests under normal ambient conditions of temperature,
humidity and pressure, with line voltage.
The leakage currents correspond to 110% of rated voltage for the tested
unit. If this is not taken into account when performing leakage tests, the
measured values must be calculated.
Safety tester for measurements according to 60601-1.
After initial setup.
Once every 12 months, thereafter, as part of a regular maintenance
plan.
Whenever power sources are affected by servicing.
Connect the power cord to the ACDC adaptor and adaptor out put to the
unit. Check for the resistance between power cord ground and printer
metal base. If the value of resistance is greater than 1 Ohm, then it
results in failure.
NOTE
The MAC600 is categorized as a Class I device.
NOTE
All shipments of MAC 600 devices to China are Class I. However,
for all other countries, the MAC 600 installed base is Class II if
the serial number is less than SF713500001PA.
The following electrical diagram shows the Measuring Circuit [M]
required for leakage current. The reading in mV corresponds to mA
(leakage current). The safety tester instruments generally work with this
TM
MAC
600
2047426-002
Revision G