Ventilation modes
Changing the Trigger Type
For double limb circuits, changing Trigger type can be done while the device is ventilating or in Standby.
To change between pressure and flow trigger types on double circuits:
1. From the Setup menu, select Circuit.
2. From the Trigger type option, press Change. The current Trigger Type is highlighted.
3. Select Flow. You will be returned to the Circuit screen where the changed Trigger type will be
displayed.
92