Using the Astral device
To change programs:
1. Select P2: (A)CV. P2 becomes the active program.
2. Select and start the Setup assistant. Refer to Using the Setup Assistant (see page 16).
Once the Setup assistant is complete, P2 will now be configured to the settings you have chosen
and is ready to Start ventilation. In the example below, P2 has been changed from (A)CV mode to
P(A)CV.
Note: Each program retains its own circuit configuration. When switching between programs, ensure you use the
correct circuit as 'learned' for that program.

Setting up a new program

Adjustments to program settings can only be made to the current active program. To make changes to
settings of other programs, you must first activate that program.
Adjustments to patient settings can be made while the device is Ventilating or in Standby mode.
WARNING
Always review and adjust alarm Settings when setting up a new program. For information, refer to
Adjusting alarm settings (see page 108).