ResMed Astral Series Clinical Manual page 178

Hide thumbs Also See for Astral Series:
Table of Contents

Advertisement

Parameter
PS
(cmH
O)
2
Min PS
(cmH
O)
2
Max PS
(cmH
O)
2
IPAP
(cmH
O)
2
Vt
(mL)
PIF
(L/min)
Ti
(sec)
Valve Ventilation and Mouthpiece
Circuit
Appendix B: Ventilation parameters
Leak Ventilation
PS
Spontaneous breaths: V-SIMV,
P-SIMV
0 to 50 [2]
AutoEPAP Off: 0 to 50 [20]
AutoEPAP On: 8 to 50 [20]
Adult: 4 to 50 [12]
Paed: 4 to 50 [12]
V-SIMV (mandatory breaths),
(A)CV
When Volume Breath option is
set to PIF:
(A)CV, V-SIMV (mandatory
breaths)
When Volume Breath option is
set to Ti:
(A)CV, V-SIMV (mandatory
breaths)
P(A)CV, P-SIMV (mandatory
breaths)
P(A)C
Setting
Adult: 2 to 50 [7]
Paed: 2 to 50 [7]
Adult: 2 to 50 [7]
Paed: 2 to 50 [7]
Adult: 100 to 2,500 [500]
Paed: 50 to 300 [100]*
Adult: 10 to 120 [50]
Paed: 5 to 60 [10]
Adult: 0.3 to 3.0 [1.0]
Paed: 0.3 to 3.0 [0.6]
Adult: 0.2 to 5.0 [1.0]
Paed: 0.2 to 5.0 [0.6]
Adult: 0.3 to 4.0 [1.0]
Paed: 0.3 to 4.0 [0.6]