Appendix B: Ventilation parameters
Features
Apnoea Vt
(mL)
Apnoea Flow Shape
(%)
Apnoea PIF
(L/min)
Apnoea P Control
(cm H
O)
2
Safety Vt (mL)
Max PS
P control max (cm H
O)
2
IPAP Max
*Adult Tapnoea can be extended to 15min when the mouthpiece interface is selected.
176
Valve Ventilation and Mouthpiece
Circuit
Leak Ventilation
Setting
If Apnoea Response is (A)CV +
Alarm:
Adult: 100 to 2500 [500]
Paed: 50 to 500 [100]
Constant
When the Volume Breath is set
to PIF.
If Apnoea Response is (A)CV +
Alarm:
Adult: 10 to 120 [50]
Paed: 5 to 60 [10]
If Apnoea Response is P(A)CV +
Alarm:
Adult: 2 to 50 [7]
Paed: 2 to 50 [7]
Adult: Off, 100 to 2500 [Off]
Paed: Off, 50 to 500 [Off]
PS setting to 50 [PS +5]
P control to 50 [P control +5]
IPAP to 50 [IPAP +5]