ResMed Astral Series Clinical Manual page 80

Hide thumbs Also See for Astral Series:
Table of Contents

Advertisement

Parameter
Resp.rate (per min)
PEEP (cmH
O)
2
PS (cmH
O)
2
Vt (Tidal Volume) (mL)
PIF (L/min)
Ti (Inspiratory time) (sec)
Flow shape (%)
Cycle (%)
Trigger type
Trigger
Rise Time (msec)
*The International ventilator standard indicates that Paediatric patient type is intended to be used for a patient receiving less
than 300 mL, however Astral permits adjustment of 'Vt' setting parameter up to 500 mL for cases where 'Vt' is set such that it
compensates for leak in the breathing circuit.
WARNING
ResMed does not recommend 500 mL as the upper limit for the pediatric tidal volume use;
however, clinicians may choose this upper limit based on their clinical determination.
Setting
Mandatory breaths:
Adult: 2 to 50 [15]
Paed: 5 to 80 [15]
Off, 3.0 to 20.0 [5.0]
Spontaneous breaths:
Adult: 2 to 50 [7]
Paed: 2 to 50 [7]
Mandatory breaths:
Adult: 100 to 2,500 [500]
Paed:50 to 300 [100]*
When Volume Breath option is set to PIF
Mandatory breaths:
Adult: 10 to 120 [50]
Paed: 5 to 60 [10]
When Volume Breath option is set to Ti
Mandatory breaths:
Adult: 0.3 to 3.0 [1.0]
Paed: 0.3 to 3.0 [0.6]
Mandatory breaths:
100 (Constant), 75, 50, 25 [100]
Spontaneous breaths:
5 to 90, Auto [Auto]
Flow/Pressure
When Trigger Type is set to Flow:
(double limb circuit only)
Adult: 0.5 to 15 [1.0] (L/min)
Paed: 0.5 to 15 [0.5] (L/min)
When Trigger Type is set to Pressure:
(double and single limb circuit)
Very Low to Very High [Medium]
Spontaneous breaths:
Min, 150 to 900 [200]
Ventilation modes
English
73