Oximetry Alarms; Spo2; Pulse Rate; Apnoea Alarm - ResMed Astral Series Clinical Manual

Hide thumbs Also See for Astral Series:
Table of Contents

Advertisement

Oximetry alarms

The oximetry alarms are only available when the pulse oximeter is connected to the Astral device. The
alarms are enabled even when the ventilator is in standby.
The SpO
and Pulse rate alarms are automatically disabled when the pulse oximeter is disconnected
2
from the Astral device.

SpO2

Oxygen saturation alarms activate when the monitored SpO
alarm setting (no delay).
Alarm
Activates when
Low SpO
Saturation of peripheral oxygen is less than the low SpO
2
measured by the pulse oximeter.
High SpO
Saturation of peripheral oxygen is greater than the high SpO
2
setting as measured by the pulse oximeter.
No SpO
monitoring SpO
2
2
disabled or disconnected, or has a degraded signal for more than 10
seconds.

Pulse rate

Pulse rate alarms activate when the average pulse rate (Pulse) exceeds the corresponding alarm setting
(no delay).
Alarm
Activates when
Low pulse rate
Pulse rate is less than the low Pulse setting as measured by the
pulse oximeter.
High pulse rate
Pulse rate is greater than the high Pulse setting as measured by the
pulse oximeter.

Apnoea alarm

For details on how to configure the Apnoea alarm and activation conditions refer to Apnoea settings (see
page 97).

Ventilation stopped alarm

The ventilation stopped alarm alerts the carer that a user has stopped the ventilator.
This alarm should be enabled if unexpected user intervention is possible.
Alarm
Activates when
Ventilation stopped
The Stop Vent. button is pressed and Stop Vent. prompt is
confirmed.
alarm settings are on and the pulse oximeter has been
Alarm settings and conditions
parameter exceeds the corresponding
2
Alarm setting (SpO
setting as
Off, 50 to 99 [85]
2
Off, 51 to 100 [Off]
2
N/A
Alarm setting (per/min)
Off, 20 to 249 [30]
Off, 21 to 250 [150]
Alarm setting
Off / On [Off]
%)
2
English
119

Advertisement

Table of Contents
loading

This manual is also suitable for:

Astral 100Astral 150

Table of Contents