Parameter
Apnoea PIF
Apnoea P control
Manual Breath Magnitude
Sigh Interval
Sigh Magnitude
* When Resp. rate ≥ 8/min and Pressure ≤ 30 hPa.
**The International ventilator standard indicates that Paediatric patient type is intended to be used for a patient receiving less
than 300 mL, however Astral permits adjustment of 'Vt' setting parameter up to 500 mL for cases where 'Vt' is set such that it
compensates for leak in the breathing circuit.
WARNING
ResMed does not recommend 500 mL as the upper limit for the pediatric tidal volume use;
however, clinicians may choose this upper limit based on their clinical determination.
Notes:
All flows and volumes are measured at BTPS conditions.
To achieve specified accuracies, a successful Learn Circuit must be performed prior to measurement testing.
Delivered pressure is regulated at the patient port.
Range
When Volume Breath is set
to PIF.
If Apnoea Response is
(A)CV + Alarm:
Adult: 10 to 120 L/min
Paed: 5 to 60 L/min
When Apnoea Response is
P(A)CV + Alarm:
2 to 50 hPa
100 to 250%
3 to 60 min
120 to 250%
Resolution
Accuracy
1 L/min
N/A
0.2 hPa
±(0.5 hPa+ 5% of target)
10%
N/A
1 min
N/A
10%
N/A
Accuracy of controls