Circuits and patient interfaces

The following circuits are available for use with the Astral device.
Single limb circuit with intentional leak (provides a leak compensated estimate of patient flow and
exhaled tidal volume)
Single limb circuit with expiratory valve
Double limb circuit (exhaled gas returns to the device for monitoring) (Astral 150 only)
Mouthpiece circuit (tube-only circuit for use with mouthpiece interfaces)
The Astral device is compatible with the following invasive and non-invasive patient interfaces:
Invasive
endotracheal tubes
tracheostomy tubes
Non-invasive
vented and non-vented masks
mouthpiece.
WARNING
The measurement of patient exhaled gas volume may be affected by leak.

Patient circuit accessories

The following patient circuit accessories are available for use with the Astral device:
Humidifier
Heat Moisture Exchange (HME)
Antibacterial filter
For information on attaching and using patient circuit accessories, refer to Attaching patient circuit
accessories (see page 48).

Power accessories

WARNING
The Astral device should only be used with accessories recommended by ResMed. Connection
of other accessories could result in patient injury or damage to the device.
Connecting the Astral device to the battery of a battery-powered wheelchair may affect the
device performance and may result in patient harm.
The Astral device can be connected to a range of accessories as follows:
Astral External Battery
ResMed Power Station II
Astral DC adapter
ResMed Remote Alarm II
Pulse Oximeter.