Disabling (or enabling) the Disconnection Alarm

The Disconnection Alarm default setting is ON. To disable the Disconnection Alarm, the patient interface
setting must be set to:
Mouthpiece (for valved circuits), or
Mask (for leak circuits).
To disable or enable the Disconnection Alarm function:
1. Select the Disconnect Alarm tab.
2. Press the slider to turn ON/OFF. A warning message will be displayed. The slider will not be shown
for invasive interfaces or mask with valve circuits.
3. Select Apply from the bottom bar to proceed. An alert message will be displayed.
4. Press Confirm on the bottom bar to continue.
5. Once confirmed, the Disconnect Alarm is disabled and the Disconnect Alarm OFF icon will be
displayed next to Standby on the Information bar.