List Of Display Messages - Miele PLW 6111 Operating Instructions Manual

Laboratory glassware washer
Table of Contents

Advertisement

 

14.2 List of display messages

Display
message
press start
no chemical 1
no chemical 2
no chemical 3
no chemical 4
salt loading
pump steam
cond.
pressure probe
- open door -
wait
close door!
ON PRINT
NO DISINFECTION The program was interrupted and the wash items were not disinfected.
REV.0.01_COD.650184_A4
Description
A program can be started.
Lack of product in DOS 1.
Lack of product in DOS 2.
Lack of product in DOS 3.
Lack of product in DOS 4.
Refill the reactivation salt container.
Steam condenser activated for a specified time to bring the amount of condensate
below the max. level.
Wash pump pressure monitoring fault
The door is open and not locked.
General reminder to wait before next input.
Close the door.
Appears when an attempt is made to start a program with the START button during
printing.
LIST OF WARNINGS AND EVENTS
Page 47

Hide quick links:

Advertisement

Table of Contents
loading