LIST OF WARNINGS AND EVENTS
Warning
34
check temp.
35
Serial connect.1
36
Serial connect.2
37
CANserialconnec.
39
no tank heating
40
no boilerheating
41
no boilerheating
42
magnetothermic
43
Phase sequence
44
levels steamcond
45
pump steam cond.
46
pump
47
flowmeter fail.1
48
flowmeter fail.2
49
flowmeter fail.3
50
flowmeter fail.4
51
flow.cold water
52
flow.warm water
53
flow.demi water
54
airfilter obstr.
55
conduc. probe
56
conductivity
58
no tank heating
59
no boilerheating
60
TIME
72
DRYING
73
DRYING
74
boilerprobelim°C
75
boiler probe
76
BOILER
77
BOILER
78
BOILER
REV.0.01_COD.650184_A4
Display message
Description
Temperature difference between the two sensors is
greater than 1°C (1.8°F).
No connection between main PCB and loading
controller
No connection between the expansion board and the
keyboard (unloading side).
No connection between master and slave (CANbus).
No heating in wash cabinet. No temperature increase
during allocated time.
Problem with boiler heating.
Problem with boiler heating.
Steam heating fault
Phase sequence is incorrect.
Steam condenser fault
Min. and max. level active drain pump fault.
Circulation pump on; pressure switch closed.
Circulation pump is rotating in the wrong direction.
DOS1 dispenser fault
DOS2 dispenser fault
DOS3 dispenser fault
DOS4 dispenser fault
Water intake fault (cold water).
Water intake fault (hot water).
Water intake fault (fully demineralized water).
Air filter is obstructed. Change the air filter.
Vacuum switch defective. No compressed air available
(check compressed air supply).
Conductivity sensor fault.
Conductivity exceeds maximum threshold.
No temperature increase during allocated time.
No temperature increase during allocated time.
Maximum time exceeded in main wash phase
Exhaust air pressure: fault when opening or closing
exhaust air valve.
Exhaust air pressure: exhaust air limit switch (open and
close both active at the same time).
The boiler temperature is too high.
Boiler probe 2 defective
The max. level on tank 3 is active and the min. level is
inactive.
The max. level on tank 1 is active and the min. level is
inactive.
The max. level on tank 2 is active and the min. level is
inactive.
Page 46