Miele PLW 6111 Operating Instructions Manual page 52

Laboratory glassware washer
Table of Contents

Advertisement

MAINTENANCE
Responsible
person: User
PROCEDURE:
Use a damp cloth to clean the machine casing. Use only pH-neutral detergents. Do not
use abrasive detergents, solvents, and/or thinners.
CLEANING THE CONTROL PANEL:
Clean the control panel with a damp cloth and dish detergent. Do not use abrasive
detergents, solvents, and/or thinners.
Responsible
person: User
PROCEDURE:
The machine must always be empty when descaling is carried out. Add 50 ml of
descaling agent directly to the wash cabinet. Start a cleaning program and allow to run
until completion.
REV.0.01_COD.650184_A4
CLEANING THE MACHINE CASING
Frequency: Daily
DESCALING
Only required for machines without an internal water softener.
Frequency: Weekly; daily with high water consumption or hard
water.