List Of Warnings And Events; List Of Warnings - Miele PLW 6111 Operating Instructions Manual

Laboratory glassware washer
Table of Contents

Advertisement

 

14. LIST OF WARNINGS AND EVENTS

Messages and warnings may appear in the display during operation.
A warning remains on display until the underlying cause is dealt with. A warning interrupts the
current program.

14.1 List of warnings

Warning
1
power fail
2
open load. door
4
load.door fail.
7
unblock.door 1
9
unlocking 1fail.
11
no cold water
12
no hot water
13
no demin. water
14
no c.+hot water
15
no c.+demi water
16
no h.+demi water
17
no chemical 1
18
no chemical 2
19
no chemical 3
20
no chemical 4
21
heating element1
23
drain problem
24
fan problem
25
drying min°C
26
prewash max°C
27
tank probe lim°C
28
dryingprobelim°C
29
boilerprobelim°C
30
tank probe
31
tank probe 2
32
drying probe
33
boiler probe
REV.0.01_COD.650184_A4
Display message
LIST OF WARNINGS AND EVENTS
Description
Signals a power failure during a program.
Loading door open and/or unlocked during a program.
Loading door locked but open.
Door fault.
Door release timeout
Cold water impulse reading from flow meter
Cold water impulse reading from flow meter – timeout
Cold water impulse reading from flow meter – timeout
Cold and hot water level – timeout
Cold and DI water level – timeout
Hot and DI water level – timeout
Lack of process chemicals (DOS1)
Lack of process chemicals (DOS2)
Lack of process chemicals (DOS3)
Lack of process chemicals (DOS4)
Fault with heating element 1
Minimum tank level during draining – timeout
Fan switch ON, pressure switch open
Air temperature does not reach the specified value
during drying.
Tank temperature exceeds specified value during
prewash.
Cabinet temperature exceeds 102°C (215°F).
(= max. setting 95°C + safety level 7°C)
(Tank heating switch or tank sensor defective)
Air temperature exceeds 162°C (323°F).
(= max. setting 150°C + safety level 12°C)
(Drying heating switch or air sensor defective)
Tank
1
temperature exceeds 100°C (212°F).
(= max. setting 80°C + safety level 20°C)
(Tank 1 heating switch or tank 1 sensor defective)
Cabinet temperature sensor 1 fault
Cabinet temperature sensor 2 fault
Drying temperature probe fault
Boiler temperature probe fault
Page 45