Agilent Technologies 1290 Infinity User Manual page 66

Binary pump
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

4
Using the Pump
Setting up the Pump with the Instrument Control Interface
Table 4
Method parameters
Parameter
Limits
Pressure Limits
Max: 1200 bar
(17400 psi ) for flow
rates up to
2 mL/min. For flow
rates between
2 mL/min and
5 mL/min, the
maximum pressure
ramps down to
800 bar (11600 psi).
Min: any value
between 0 and the
upper pressure limit
setting.
Timetable
Advanced
External Contacts
66
Description
Sets the maximum and minimum pressure limits for the pump.
• Max is the maximum pressure limit at which the pump will switch
itself off, protecting the analytical system against over-pressure.
• Min is the minimum limit at which the pump will switch itself off, for
example, if any solvent reservoir is empty, this prevents system
damage by pumping air.
See
"Timetable Settings"
See
"Advanced Settings"
The External Contacts section enables you to set up the switching of the
external contacts.
N O T E
The External Contacts section is present only when a
BCD/external contacts board is installed.
Agilent 1290 Infinity Binary Pump User Manual
on page 68
on page 67